| Model Number |
CI-1400-01 |
| Medical Device Problem Codes |
Electrical /Electronic Property Problem (1198); Electrostatic Discharge (2149); Intermittent Communication Failure (4038)
|
| Health Effect - Clinical Code |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
09/12/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
| |
|
Event or Problem Description
|
|
The recipient is reportedly experiencing loss of lock.External equipment was exchanged; however, the issue was not resolved.Integrity testing could not be completed due to no lock.Revision surgery is under consideration.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record was performed at the request of the nca and no anomalies were noted.Revision surgery is scheduled.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
| |
|
Additional Manufacturer Narrative
|
|
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: sections b.3 & d.6b.The recipient's device was explanted.The recipient was reimplanted with another advanced bionics cochlear device.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
| |
|
Additional Manufacturer Narrative
|
|
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
| |
|
Additional Manufacturer Narrative
|
|
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed sliced silicone overmold on the top and bottom covers, as well as a severed electrode.This is believed to have occurred during revision surgery.The device passed photographic imaging inspection.System lock could not be obtained at any spacing.The electrode condition prevented an electrical test from being performed.The device passed some of the electrical tests performed.The device passed some the mechanical tests performed.The failure of this device is attributed to a short from a power node to the case ground at the analog chip.It is believed that electrostatic discharge (esd) led to an overload voltage, damaging structures on case ground node inside the analog integrated circuit.This ultimately caused the device to cease functioning.A corrective action was implemented.This older device configuration is not currently manufactured.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
| |
|