• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS IGM-2; IGM (MU CHAIN SPECIFIC)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS IGM-2; IGM (MU CHAIN SPECIFIC) Back to Search Results
Catalog Number 05220726190
Medical Device Problem Code Non Reproducible Results (4029)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/01/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The serial number of the c 702 analyzer is (b)(6).The investigation is ongoing.
 
Event or Problem Description
The initial reporter stated they received discrepant results for one patient sample tested with igm-2 on a cobas 8000 c702 module.The sample initially resulted in an igm value of 39.66 g/l with a data flag.The sample was repeated three additional times using a decreased sample volume, resulting in igm values of 70.63 g/l with a data flag, 69.39 g/l, and 69.93 g/l.The sample was diluted 1:20 and repeated, resulting in an igm value of 36.93 g/l with a data flag.The sample was repeated using an increased sample volume, resulting in an igm value of 0.14 g/l.
 
Additional Manufacturer Narrative
High igm values are plausible and are due to the patient's condition (waldenstrom¿s macroglobulinaemia / lymphoplasmacytic lymphoma).The very low value of 0.14 g/l obtained using increased sample volume points to a high dose hook effect.For highly concentrated samples, dilutions using decreased sample volume or manual off-board dilutions are recommended.The 1:20 dilution was performed on board.Ordering additional dilutions directly on the instrument will cancel the automatic pre-dilution of the test and can potentially give false results due to missing pre-dilution.The investigation did not identify a product problem.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IGM-2
Common Device Name
IGM (MU CHAIN SPECIFIC)
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhofer strasse 116
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
indianapolis, IN 46250
MDR Report Key17828838
Report Number1823260-2023-03122
Device Sequence Number7182344
Product Code DAO
UDI-Device Identifier04015630925742
UDI-Public04015630925742
Combination Product (Y/N)Y
Initial Reporter CountryUK
PMA/510(K) Number
K040431
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number05220726190
Device Lot Number716687
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/04/2023
Supplement Date Received by Manufacturer11/23/2023
Initial Report FDA Received Date09/27/2023
Supplement Report FDA Received Date12/08/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CHEMOTHERAPY.
Patient Age66 YR
Patient SexMale
-
-