|
A visual inspection was performed on the returned device.The reported separation was confirmed.The reported difficulty advancing and removing the device from the introducer sheath could not be replicated in a testing environment due to the condition of the returned device.The reported difficulty advancing and removing the device from the anatomy could not be replicated in a testing environment as it was based on operational circumstances.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.As difficulty removing the device was experienced, the account reported that force was applied, and a separation occurred.It should be noted that the coronary dilatation catheters (cdc), nc trek rx, global, instruction for use (ifu) states: if resistance is felt, determine the cause before proceeding.Continuing to advance or retract the catheter while under resistance may result in damage to the vessels and / or damage / separation of the catheter.In this case, it is likely that the violation of the ifu contributed to the reported separation as the device separated after force was applied.The investigation was unable to determine a conclusive cause for the reported difficulty advancing and removing the device from the introducer sheath.The reported difficulty advancing and removing the device from the anatomy and unexpected medical intervention appear to be related to circumstances of the procedure.Additionally, the reported shaft separation appears to be related to user error/operational context given the clinical situation.Possible factors which may contribute to resistance with the introducer sheath during advancement include, but are not limited to, manufacturing, damage to the introducer sheath, or procedural technique (devices not properly supported or coaxially aligned).Possible factors that can contribute to difficult to remove/resistance with the introducer sheath post-inflation include, but are not limited to, loose balloon folds, introducer sheath lumen obstructions, kinks, bends or procedural technique.Furthermore, it was reported that the bdc interacted with the moderately calcified and mildly tortuous anatomy resulting in the reported difficulty advancing and removing the device from the anatomy.Force was applied in an attempt to remove the device, as difficulty was encountered, and the shaft ultimately separated.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
|