• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ELLIPSE DR ICD; Pulse generator, permanent, implantable

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ELLIPSE DR ICD; Pulse generator, permanent, implantable Back to Search Results
Model Number CD2311-36
Medical Device Problem Code Defective Device (2588)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 09/14/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No complaint was received with the return of the device.Failure event observed during analysis.Final analysis found premature battery depletion.Interrogation of the device revealed the device was at elective replacement indicator (eri) when received.A longevity calculation was performed and found the battery depletion was premature based on the device usage.Electrical testing revealed high current drain from the hybrid.A hybrid anomaly was found to be the cause of the high current drain and premature battery depletion.
 
Event or Problem Description
This report is to advise of an event observed during analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELLIPSE DR ICD
Common Device Name
Pulse generator, permanent, implantable
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key17831143
Report Number2017865-2023-47160
Device Sequence Number3389994
Product Code NVZ
UDI-Device Identifier05414734505987
UDI-Public05414734505987
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P910023
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 09/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date04/30/2014
Device Model NumberCD2311-36
Device Lot Number3858236
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/11/2019
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/14/2023
Initial Report FDA Received Date09/28/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/23/2012
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-