• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AIZU OLYMPUS CO., LTD. CONNECTING TUBE; ENDOSCOPE REPROCESSOR ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIZU OLYMPUS CO., LTD. CONNECTING TUBE; ENDOSCOPE REPROCESSOR ACCESSORIES Back to Search Results
Model Number MAJ-2114
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/05/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is ongoing and follow up with the user facility is currently being performed.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Event or Problem Description
The olympus representative reported (on behalf of the customer) that the connecting tube wing broke after a few uses.The issue was found in the in cleaning and disinfection room.There was no patient involvement and no user injury.Related patient identifiers: (b)(6).
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.The device history record was unable to be reviewed for this device since the serial number was not provided.However, olympus only releases products to market that meet all manufacturing specifications and final product release criteria.Based on the results of the investigation, the connection tube was damaged.However, the definitive root cause of the broken connecting tube wing could not be determined.Olympus will continue to monitor field performance for this device.It is possible that the connecting tube was broken by external stress on its lock lever through the application of force in the incorrect direction.The instruction manual identifies the following verbiage, which may have prevented the phenomenon: ¿preparation and inspection: visually inspect the connecting tube to ensure that there are no cracks, breaks, rips, scratches, attached debris, or other irregularities.Move the lock levers of the reprocessor side connector to make sure that they function properly and are not broken.¿ olympus will continue to monitor field performance for this device.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation since they received the device.The customer's complaint was confirmed.Visual inspection was performed on the as is received condition of the connecting tube; noted that the connecting tube was returned without the original packaging.Both lock levers are broken.The broken portions were not returned for evaluation.The pin inside the connecting tube is present and looks to be in good condition.The tubing was inspected and there are no signs of punctures on the tubing or signs of residue inside the tubing.Based on the results of the investigation, it is likely that the connecting tube was broken by external stress on its lock lever, and could not be connected.As a cause of the external stress, it is likely that the user applied force in incorrect direction.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONNECTING TUBE
Common Device Name
ENDOSCOPE REPROCESSOR ACCESSORIES
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key17833293
Report Number9610595-2023-14218
Device Sequence Number17865060
Product Code FEB
UDI-Device Identifier04953170404092
UDI-Public04953170404092
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K190969
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMAJ-2114
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/20/2023
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/05/2023
Supplement Date Received by Manufacturer10/26/2023
02/15/2024
Initial Report FDA Received Date09/28/2023
Supplement Report FDA Received Date10/26/2023
02/19/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-