MENTOR TEXAS MENTOR SMOOTH ROUND SPECTRUM; PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE
|
Back to Search Results |
|
| Catalog Number |
3501480 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Deformity/ Disfigurement (2360)
|
| Date of Event |
01/01/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a female who underwent a breast augmentation revision with a mentor smooth round spectrum, 475cc saline breast implant experienced left-sided deflation post procedure.The issue was diagnosed through in office examination.As a result, patient underwent removal on (b)(6) 2023.
|
| |
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.Since no lot number was provided, no manufacturing record evaluation could be performed.Reason for device explant and/or reoperation: deflation.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Suspect device received on october 03, 2023 device evaluation completed on october 04 , 2023: the product was returned to mentor for evaluation.Mentor conducted visual inspection and leak testing of the returned device.Visual analysis of the returned sample revealed that the spec smooth rnd 475cc breast implant had a crease/fold on the anterior view.Leak testing was performed, according to mentor procedure, and it identified a tear within a crease/fold measuring approximately 0.1 cm.The evaluation determined that the possible cause of the rupture is consistent with normal wear.Deflation can occur at any time after implantation, but it is more likely to occur the longer the implant is implanted.Creating wrinkles or folds should be avoided during implantation or other procedures as it can result in a deflation in a later time.Breast implants may also simply wear out over time.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Per additional information received on (b)(6) 2023, patient underwent unilateral replacement with l cat:3502400 sn :(b)(6).Manufacturer¿s reference number: (b)(4).
|
| |
|