DEPUY IRELAND - 9616671 ATTUNE SHIM SZ7 6MM; TIBIAL INSERT TRIALS Back to Search Results
Catalog Number 254500672
Medical Device Problem Codes Break (1069); Improper or Incorrect Procedure or Method (2017)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/19/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that the instrument broke while trailing due to incorrect use.Patient impact ¿ delay in surgery as we had to wait to open another instrument.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: an analysis of the product could not be performed since a physical sample was not received for evaluation.An evaluation of the manufacturing record could not be performed as the required product identification number was not provided to complete the evaluation.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.However, if the product or product id numbers are received at a later date, the investigation will be updated as applicable.
 
Event or Problem Description
Additional information received.A.What part of the instrument was broke? -the shim snapped into two pieces, it¿s the entire instrument, see attached customer photo.B.Did it break into two or more pieces? if no, please specify what is the alleged deficiency, is it bent, cracked, stripped, scratched, worn, cross threaded or any device interaction issues? -its snapped in two.C.It was indicated that there was a surgical delay, what is the duration of the delay? -5minutes.D.Was there any adverse consequences that affected the patient because of the reported event? -no.E.Can you please clarify how the device was used incorrectly? -do much force when extracting the trial insert.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE SHIM SZ7 6MM
Common Device Name
TIBIAL INSERT TRIALS
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthpaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key17835487
Report Number1818910-2023-19679
Device Sequence Number7695378
Product Code HWT
UDI-Device Identifier10603295133544
UDI-Public10603295133544
Combination Product (Y/N)N
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number254500672
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/18/2023
Supplement Date Received by Manufacturer10/04/2023
Initial Report FDA Received Date09/28/2023
Supplement Report FDA Received Date10/10/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-