Blank fields on this form indicate the information is unknown or unavailable.E1- customer (person) : (b)(6).E3- occupation: technical support engineer.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned or that a death or serious injury occurred; nor is it admission that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|
Blank fields on this form indicate the information is unknown, unavailable, or unchanged.Additional information: b5 event summary it was reported that the green laser light was visible through the blue coating of a cook® multi-use holmium laser fiber.The issue "was along the entire length of the fiber".The fiber was a new fiber and had not been used yet.The procedure was completed with a new fiber.There were no adverse effects reported.Investigation ¿ evaluation reviews of the complaint history, device history record (dhr), instructions for use (ifu), manufacturing instructions (mi), and quality control procedures, as well as a visual inspection of the returned device, were conducted during the investigation.One device was returned to cook in opened packaging.It was missing the clear tip, and the blue coating has a scorched appearance.A document-based investigation evaluation was performed.A review of the device master record (dmr) concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.A review of the dhr for the reported complaint device lot revealed no recorded non-conformances relevant to the failure mode.A database search did not identify any other related complaints with the reported device lot.Because adequate inspection activities have been established, there is objective evidence that the dhr was fully executed and no other lot related complaint has been received from the field, it was concluded that there is no evidence that nonconforming product exists in house or in field.Cook also reviewed product labeling.The product ifu, [t_h30m_rev2] ¿h-30 holmium laser fiber (multi-use),¿ provides the following information to the user related to the reported failure mode: warnings ¿ do not bend fiber at sharp angles.If visible light (aiming beam) can be seen leaking from the fiber, fiber failure may result when therapeutic energy is applied as the fiber is deflected beyond the optical limits of total internal reflection.Discard fiber if visible light leakage is observed.¿ any deviation(s) to the reprocessing instructions prescribed in this ifu may damage the device and result in a device which has compromised sterility or is not fully reprocessed to relevant standards and guidelines.Precautions ¿ to avoid reducing the life of the fiber, follow these precautions: ¿ do not improperly handle the fiber.¿ do not lase at high power for extended periods of time.¿ do not lase continuously with the fiber tip in contact with the tissue.¿ do not lase with a fiber that has a contaminated or damaged proximal end.¿ do not operate a laser with poor beam alignment or focus.¿ do not subject the fiber to sharp bends during handling, use, or storage.¿ always keep the proximal connector end dry and free of contaminants.¿ discard any laser fiber that is cracked or broken or does not meet minimum power transmission standards.¿ do not use this laser fiber in the presence of flammable anesthetics or combustible materials.¿ ferrule dust cap should stay attached when fiber is not connected to the laser system.¿ do not exceed recommended power limits.Instructions for use 9.Check the integrity of laser fiber by using visible aiming beam from laser system.Observe visible light profile out of laser fiber tip.The profile should be a well-defined round or oval pattern.If output beam is not round/oval pattern, or visible light leaks from any spot along the length of the fiber shaft, then the fiber core is broken and the fiber should be discarded as hazardous waste.Based on the information provided, inspection of the returned device, and the results of the investigation, a definitive cause of the laser fiber breakage could not be determined.Many factors could have contributed to the laser fiber breakage such as device deflection, variation in the material properties, or having an energy applied that exceeds the recommended power limits.Cook will continue monitoring of similar complaints and has notified the appropriate personnel of this event.¿ per a review of risk documentation, no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|
Blank fields on this form indicate the information is unknown or unavailable.Correction: d9, h3 h3: device evaluated by mfg = other (81) - device evaluation has begun; however, a conclusion is not yet available.Device has been returned and preliminary evaluation has been performed, however, our investigation is ongoing and device evaluation summary will be included in our follow up report once our investigation has been completed.This report includes information known at this time.Follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|