|
It was reported that (b)(6) 2023 the surgeon had issues with screws getting purchase in the bone.It's possible that there was a fragment there, however it was unknown.He decided to leave the third screw out because he felt comfortable with the fixation with two screws.Additionally when the first 1.1 drill bit was selected before use, it was determined to be bent.It was removed, and another drill bit was used successfully.The surgery was delayed for 5 minutes but it was successfully completed.No patient consequences was there.This report is for one (1) 1.5 cortex screw slf-tpng t4 sd rec 14.This is report 1 of 3 for complaint (b)(4).
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: d2b: additional product code: hwc.D9: complainant part is not expected to be returned for manufacturer review/investigation.H3, h4, h6: without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|