• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DORNOCH DUO FLUID CART WITH POWER IV, REFURBISHED; APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DORNOCH DUO FLUID CART WITH POWER IV, REFURBISHED; APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED Back to Search Results
Model Number N/A
Medical Device Problem Codes Thermal Decomposition of Device (1071); Failure to Power Up (1476)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/05/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An investigation into the reported event has been initiated under (b)(4).Once the investigation has been completed, a follow up report will be submitted with the investigation results and any actions taken by the manufacturer.
 
Event or Problem Description
It was reported that outside of surgery, the unit had no power, and the ac cord has burnt mark.The power inlet module was not damaged.The unit had physical evidence of burn.The account did not have the cart plugged into a power strip or extension cord.There was no patient involvement.Due diligence is complete, no further information is available.
 
Additional Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).Review of the most recent repair record determined the unit was confirmed to not power on.The power inlet module was found to have melted and showed signs of bubbling and the power cord was burnt.The power inlet module was replaced and resolved the reported issue.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.The root cause of the reported issue is attributed to a component failure.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.Upon review of our reporting decision on blown fuses exhibiting signs of overheating.A review of all complaints for this product indicates that there has never been an event that has resulted in a death, serious injury, and/or actual fire when the fuse has blown or fuse drawer deformation.Therefore, zimmer biomet will not be reporting this failure mode going forward.
 
Event or Problem Description
There is no additional information available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DUO FLUID CART WITH POWER IV, REFURBISHED
Common Device Name
APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED
Manufacturer (Section D)
DORNOCH
200 northwest parkway
riverside MO 64150
Manufacturer (Section G)
DORNOCH
200 northwest parkway
riverside MO 64150
Manufacturer Contact
jennifer rapsavage
56 e. bell drive
warsaw, IN 46582
5745260384
MDR Report Key17837063
Report Number0001954182-2023-00097
Device Sequence Number20757061
Product Code JCX
UDI-Device Identifier00889024501645
UDI-Public(01)00889024501645(11)230420
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K190789
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue NumberUE514010300
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/05/2023
Supplement Date Received by Manufacturer01/10/2024
Initial Report FDA Received Date09/28/2023
Supplement Report FDA Received Date01/11/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Patient SexPrefer Not To Disclose
-
-