| Model Number |
BL2282 |
| Medical Device Problem Code |
Pressure Problem (3012)
|
| Health Effect - Clinical Codes |
Eye Injury (1845); Intraocular Pressure Increased (1937); Prolapse (2475)
|
| Date of Event |
09/01/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A field service engineer was dispatched to the user facility to evaluate the system.He was unable to find any faults.All function¿s of the system performed as they should.The device history was reviewed and showed the module met all specifications on the date of manufacture.The investigation is underway.
|
| |
|
Event or Problem Description
|
|
It was reported by the user facility in the united kingdom that during the case the eye became hard due to high pressure causing the iris to prolapse.When checked the infusion pressure was set to 35mmhg.The iris was redeposited successfully and the anterior chamber reformed.Surgery was completed safely, but the iop fluctuations did compromise the corneal view to some extent making the later stages of the membrane peel more challenging.I wasn¿t timing the surgery, but the vitrectomy cutter is inefficient and the gel trimming took longer than it should have.A small volume of anesthesia top-up was needed.Apart from a slightly dilated pupil, even following miochol-e, the eye is quiet with a normal iop.
|
| |
|
Additional Manufacturer Narrative
|
|
The lot history, trend analysis, risk analysis and directions for use review were considered acceptable, with the product performing within anticipated rates.No corrective action is required.Product evaluation did not confirm the failure and we are unable to determine the root cause.No corrective action required.
|
| |
|