|
This follow-up report is being submitted to relay additional information.The following sections were updated: b4, b5, d4, d9, g3, g6, h2, h3, h4, h6, h10 visual examination of the returned product identified the part received in in 2 pieces.The part is broken at oval hole.No other damage noticed.Fracture surface analysis of the dvr anatomic ext right plate sample showed that it suspected to have fractured due to fatigue.Review of the device history record(s) identified no deviations or anomalies during manufacturing.A definitive root cause cannot be determined.The reported event is confirmed as product was returned.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|