| Model Number |
BIO-COMP INTERFSCRW,FULL THD |
| Medical Device Problem Codes |
Defective Device (2588); Material Twisted/Bent (2981)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/31/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during an anterior cruciate ligament surgery the thread on the screw has twisted off.There was no harm for patient, operator or third party.The surgery was finished successfully with a new device (ar-1380tc).It was not necessary to switch the surgical technique or do a second surgery.
|
| |
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d9, g3, h6.-complaint allegation is not confirmed.-one unpackaged ar-1370tc full thread cannulated biocomposite interference screw was received for evaluation.-visual inspection noted no problems with the device.The device was further visually inspected under a magnifier and no damage was observed to the thread of the device.No problem found.-functional testing was performed with an ar-1996cd biocomposite interference screwdriver batch number: (b)(4) and it was observed that the ar-1370tc full thread cannulated biocomposite interference screw sat flush against the driver.No problem found.-no problem found.-refer to investigation photos.
|
| |
|