• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LEMAITRE VASCULAR, INC. SYNTEL SILICONE EMBOLECTOMY CATHETER - REGULAR TIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LEMAITRE VASCULAR, INC. SYNTEL SILICONE EMBOLECTOMY CATHETER - REGULAR TIP Back to Search Results
Catalog Number E1801-38
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/28/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The reported problem was confirmed.The balloon was observed to be ruptured.While the reported event was confirmed, the exact root cause of the balloon rupture could not be determined.Due to the nature of the product, balloon ruptures are an expected risk when using this product.As stated in the ifu: as with all catheterization procedures, complications may occur.These may include but are not limited to: infection, local hematomas, intimal disruption, arterial dissection, perforation and rupture, hemorrhage, arterial thrombosis, distal emboli of blood clots or arteriosclerotic plaque, air embolus, aneurysms, arterial spasms, arteriovenous fistula formation, balloon rupture or tip separation with fragmentation and distal embolization.The possibility of balloon rupture must be taken into account when considering the risks involved in the catherization procedure.The lot history record for the device was reviewed, no issues were found during manufacturing or packaging that would cause or contribute to the reported event.
 
Event or Problem Description
It was reported syntel silicone embolectomy catheter (regular tip) balloon ruptured while in use on a patient during a procedure.A replacement was used.No injury reported to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNTEL SILICONE EMBOLECTOMY CATHETER - REGULAR TIP
Common Device Name
EMBOLECTOMY CATHETER
Manufacturer (Section D)
LEMAITRE VASCULAR, INC.
63 second ave
burlington MA 01803
Manufacturer (Section G)
LEMAITRE VASCULAR, INC
63 second ave
burlington MA 01803
Manufacturer Contact
peter song
63 second ave
burlington, MA 01803
7814251683
MDR Report Key17843216
Report Number1220948-2023-00171
Device Sequence Number7381739
Product Code DXE
UDI-Device Identifier00840663101023
UDI-Public(01)00840663101023
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K992368
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Risk Manager
Type of Report Initial
Report Date (Section B) 09/29/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberE1801-38
Device Lot NumberNSE2586
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/06/2023
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/30/2023
Initial Report FDA Received Date09/29/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/16/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-