Event or Problem Description
|
|
It was reported by brazil that during service and evaluation, it was determined that the quick coupling device was frozen/would not move- chuck seized.It was further observed that the k-wire settings could not be changed, and the device would not hold the k-wire.It was further determined that the device failed pretest for check function of adjusting sleeve, check function of tension lever, check the tool coupling and check function in running mode.It was noted in the service order that during pre-surgery, it was discovered that device had a faulty trigger wire system and was unable to lock the k-wire device.It was not reported if there were any delays in the surgical procedure or if a spare device was available for use.There was no patient involvement reported.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.H10 additional narrative: d10: concomitant med products and therapy dates: k-wire device, (b)(6) 2023 device evaluation: the actual device was returned for evaluation.During repair, an evaluation was performed and it was determined that the device being frozen/would not move-chuck seized, identified during service and evaluation was confirmed.The assignable root cause was determined to be traced to user, which is use error.Udi:(b)(4).
|
| |
|
Search Alerts/Recalls
|
|
|