• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH QUICK COUPLING FOR K-WIRES; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH QUICK COUPLING FOR K-WIRES; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT Back to Search Results
Catalog Number 511.791
Medical Device Problem Codes Device-Device Incompatibility (2919); Mechanical Jam (2983)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/29/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported by brazil that during service and evaluation, it was determined that the quick coupling device was frozen/would not move- chuck seized.It was further observed that the k-wire settings could not be changed, and the device would not hold the k-wire.It was further determined that the device failed pretest for check function of adjusting sleeve, check function of tension lever, check the tool coupling and check function in running mode.It was noted in the service order that during pre-surgery, it was discovered that device had a faulty trigger wire system and was unable to lock the k-wire device.It was not reported if there were any delays in the surgical procedure or if a spare device was available for use.There was no patient involvement reported.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.H10 additional narrative: d10: concomitant med products and therapy dates: k-wire device, (b)(6) 2023 device evaluation: the actual device was returned for evaluation.During repair, an evaluation was performed and it was determined that the device being frozen/would not move-chuck seized, identified during service and evaluation was confirmed.The assignable root cause was determined to be traced to user, which is use error.Udi:(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUICK COUPLING FOR K-WIRES
Common Device Name
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer Contact
kate karberg
eimattstrasse 3
oberdorf 4436
SZ   4436
3035526892
MDR Report Key17843643
Report Number8030965-2023-12376
Device Sequence Number7173359
Product Code HWE
UDI-Device Identifier07611819156661
UDI-Public07611819156661
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 09/29/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number511.791
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/11/2023
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/20/2023
Initial Report FDA Received Date09/29/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/19/2015
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-