Model Number N/A
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/13/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).H6: proposed component (annex g) code is mechanical (g04) - provisional bottom.Reported event was confirmed by review of photograph provided.Visual evaluation of the provided device photograph shows signs of repeated use and the post feature has fractured.Device history record (dhr) was reviewed and no discrepancies relevant to the reported event were found.A previous investigation into this issue determined that the likely root cause for the tasp fractures is bending/torsional loading on the device.A notification was sent to the field on august 7th, 2014 to provide additional clarifications relating to the instructions for use of the tasp construct for all persona users on file at the time of the field notice.Ps tasp top components and standard (non-cps) tasp bottom components have been modified to prevent fracture.The fractured tasp component in this complaint was manufactured before the design modification.The root cause is considered to be a previously addressed design issue.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
It was reported that the device was fractured.Attempts to obtain additional information have been made; however, no more is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TIBIAL ARTICULAR SURFACE PROVISIONAL
Common Device Name
INSTRUMENT KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key17844068
Report Number0001822565-2023-02670
Device Sequence Number4401911
Product Code JWH
UDI-Device Identifier00889024245044
UDI-Public(01)00889024245044(10)62683418
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K113369
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/29/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number42527000707
Device Lot Number62683418
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/13/2023
Initial Report FDA Received Date09/29/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/07/2014
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-2578-2014
Patient Sequence Number1
Patient SexMale
-
-