• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SàRL STERILE LATARJET SCREW, 34MM; SCREW, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SàRL STERILE LATARJET SCREW, 34MM; SCREW, FIXATION, BONE Back to Search Results
Catalog Number 288225
Medical Device Problem Code Break (1069)
Health Effect - Clinical Codes Discomfort (2330); Unspecified Tissue Injury (4559)
Type of Reportable Event Serious Injury
Event or Problem Description
This is report 1 of 2 for (b)(4).It was reported by the sales rep in switzerland that during an unknown procedure on an unknown date, two sterile latarjet screw, 34mm devices were used.According to the report, the devices disturbed the patient and needed to be remove.It was reported that the surgeon tried to remove it with the screwdriver arthroscopically but the screws broke; and therefore, could not be removed and left the screws inside the patient.The status of the patient was unknown.No additional information was provided.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: udi: (b)(4).The lot number was unknown.D4, g1, h4: the lot number was unknown; therefore, the expiration date, manufacturing site name and device manufacture date were unknown.E3: reporter is a j&j sales representative.Investigation summary:the complaint device is not being returned, it was implanted in the patient, therefore unavailable for a physical evaluation.Since the complaint device was implanted, we cannot determine a root cause for the reported failure.If additional information is received in the future, we will reopen the complaint and perform the investigation as appropriate.Given that no lot number was provided, a manufacturing record evaluation (mre) review cannot be performed.If the lot number becomes available, the mre review will be performed.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STERILE LATARJET SCREW, 34MM
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL
chemin-blanc 38
le locle 02400
SZ  02400
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key17844269
Report Number1221934-2023-03578
Device Sequence Number4450372
Product Code HWC
UDI-Device Identifier10886705027026
UDI-Public10886705027026
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
K110763
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number288225
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/19/2023
Initial Report FDA Received Date09/29/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-