• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ADVANIX BILIARY; CATHETER, BILIARY, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ADVANIX BILIARY; CATHETER, BILIARY, DIAGNOSTIC Back to Search Results
Model Number M00534210
Medical Device Problem Codes Break (1069); Positioning Failure (1158)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/14/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported to boston scientific that an advanix biliary stent and naviflex rx delivery system was selected for use in the common hepatic duct during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6), 2023.During the procedure, the guide catheter could not be withdrawn, and the stent could not be deployed.The advanix biliary stent and naviflex rx delivery system were removed from the patient, and the procedure was completed with another advanix biliary stent.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.This event has been deemed a reportable event based on the investigation finding of guide catheter break.Please see block h10 for full investigation details.
 
Additional Manufacturer Narrative
Block h6: imdrf device code a0401 captures the reportable investigation finding of guide catheter break.Block h10: the returned advanix biliary stent with naviflex rx delivery system was analyzed, a visual evaluation noted that the guide catheter was detached from the distal section and the stent remained attached to the naviflex rx delivery system.The pull wire was found fully retracted.The reported event of stent did not deploy was confirmed.Based on all the gathered information and product analysis, the investigation concluded that most likely the reported event and observed problems were most likely due to operational factors encountered during the procedure.Stent deployment technique and/or excess of force applied during pull wire retraction may result in friction between the push catheter and guide catheter.This may have caused the detachment of the guide catheter.Therefore, a review and analysis of all available information indicated the most probable cause is adverse event related to the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANIX BILIARY
Common Device Name
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
780 brookside drive
spencer IN 47460
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key17844620
Report Number3005099803-2023-05177
Device Sequence Number14549674
Product Code FGE
UDI-Device Identifier08714729787303
UDI-Public08714729787303
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
K101314
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/29/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/21/2025
Device Model NumberM00534210
Device Catalogue Number3421
Device Lot Number0031281765
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/16/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/07/2023
Initial Report FDA Received Date09/29/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/22/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-