| Model Number |
C2201 |
| Medical Device Problem Codes |
Dull, Blunt (2407); Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/30/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Procedure performed: lavh event description: the rep was not present for the case but talked to an er tech who was in the case.Was told that when the device was initially inserted, the tip of the needle hit hard tissue and bent.The device was removed and about less than a quarter inch of the tip of the needle was bent.The case was complete with a new insufflation needle without issue.There was no patient injury and the patient is doing fine.Product is available for return intervention: the case was completed with a new one patient status: patient is fine.
|
| |
|
Additional Manufacturer Narrative
|
|
The event unit is anticipated to return to applied medical.A follow-up report will be provided following the completion of the investigation.
|
| |
|
Additional Manufacturer Narrative
|
|
The event unit was returned to applied medical for evaluation.Visual inspection confirmed that the complainant¿s experience of a bent blunt stylet.It was also observed that the stylet was unable to retract and fully expose the needle.Based on the condition of the returned unit and the description of the event, it is likely that the blunt stylet jammed and failed to expose the needle as it could not retract as intended, which resulted in the stylet bending.Correction: section device code has been updated to (b)(4).- material twisted/bent.
|
| |
|
Event or Problem Description
|
|
Procedure performed: lavh event description: the rep was not present for the case but talked to an er tech who was in the case.Was told that when the device was initially inserted, the tip of the needle hit hard tissue and bent.The device was removed and about less than a quarter inch of the tip of the needle was bent.The case was complete with a new insufflation needle without issue.There was no patient injury and the patient is doing fine.Product is available for return intervention: the case was completed with a new one patient status: patient is fine.
|
| |
|