• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION AGILE ESOPHAGEAL; PROSTHESIS, ESOPHAGEAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION AGILE ESOPHAGEAL; PROSTHESIS, ESOPHAGEAL Back to Search Results
Model Number M00517470
Medical Device Problem Codes Break (1069); Activation, Positioning or Separation Problem (2906)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/08/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Block h6: imdrf device code a15 captures the reportable event of stent partially deployed.
 
Event or Problem Description
It was reported to boston scientific corporation that an agile esophageal fully covered stent was used to treat a malignant esophageal stricture during a stent placement procedure performed on (b)(6) 2023.The patient's anatomy was not tortuous and was not dilated prior to stent placement.During the procedure, when the stent was deployed, the blue catheter was detached from the delivery system.The stent was partially deployed when it was removed from the patient.The procedure was completed with another agile esophageal stent.There were no patient complications reported as a result of this event.
 
Additional Manufacturer Narrative
Block h6: imdrf device code a15 captures the reportable event of stent partially deployed.Block h10: an agile esophageal stent and delivery system were received for analysis.The stent was received partially deployed.Visual inspection found the outer blue sheath and the outer clear sheath kinked.The outer blue sheath was detached from the delivery system.Functional inspection could not be performed due to the sheath being detached.No other problems were noted to the stent and delivery system.Product analysis confirmed the reported events of stent partially deployed and sheath break.It is most likely that procedural factors such as lesion characteristics, handling of the device, the technique used by the physician (force applied) could have contributed to the reported event of sheath break and the observed event of sheath kinked.These problems could have then resulted to the stent being unable to fully deploy during the procedure.Therefore, a review and analysis of all available information indicated the most probable cause is adverse event related to procedure.A product labeling review identified that the device was used per the instructions for use (ifu) / product label.
 
Event or Problem Description
It was reported to boston scientific corporation that an agile esophageal fully covered stent was used to treat a malignant esophageal stricture during a stent placement procedure performed on (b)(6) 2023.The patient's anatomy was not tortuous and was not dilated prior to stent placement.During the procedure, when the stent was deployed, the blue catheter was detached from the delivery system.The stent was partially deployed when it was removed from the patient.The procedure was completed with another agile esophageal stent.There were no patient complications reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AGILE ESOPHAGEAL
Common Device Name
PROSTHESIS, ESOPHAGEAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key17844902
Report Number3005099803-2023-05215
Device Sequence Number12558499
Product Code ESW
UDI-Device Identifier08714729973096
UDI-Public08714729973096
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K180144
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/17/2024
Device Model NumberM00517470
Device Catalogue Number1747
Device Lot Number0030366358
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/25/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 09/11/2023
Supplement Date Received by Manufacturer10/17/2023
Initial Report FDA Received Date09/29/2023
Supplement Report FDA Received Date11/08/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/18/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-