Model Number VTICM513.2
Medical Device Problem Codes Misfocusing (1401); Improper or Incorrect Procedure or Method (2017); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Blurred Vision (2137)
Date of Event 08/17/2023
Type of Reportable Event Serious Injury
Event or Problem Description
The reporter indicated that a 13.2mm vticm513.2 implantable collamer lens of -13.0/2.0/82 (sphere/cylinder/axis) was implanted into the patient's left (os) on (b)(6) 2023.On (b)(6) 2023, the lens was exchanged for a longer length lens due to lens rotation not associated to low vault and refractive surprise.The problem was resolved.Cause of the event was user error.
 
Additional Manufacturer Narrative
A4-a6:unk.H6: user error.Claim# (b)(4).
 
Additional Manufacturer Narrative
Corrected data: h6: 4627-explant.Claim# (b)(4).
 
Additional Manufacturer Narrative
Additional information: h3 - device evaluation: the lens was returned in liquid in a microcentrifuge vial with residue/debris on the lens.Visual inspection found the haptic broken and residue on the lens, specifically fibre.Dimensional and functional inspections found the lens to be within specifications.Claim# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Common Device Name
PHAKIC TORIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SUGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
joselene muniz
800 royal oaks dr. suite:210
monrovia, CA 91016
6263037902
MDR Report Key17846529
Report Number2023826-2023-04241
Device Sequence Number11862386
Product Code QCB
UDI-Device Identifier00840311311804
UDI-Public00840311311804
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P030016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/31/2025
Device Model NumberVTICM513.2
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/20/2023
Supplement Date Received by Manufacturer11/24/2023
12/21/2023
Initial Report FDA Received Date09/29/2023
Supplement Report FDA Received Date12/20/2023
12/26/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/29/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CARTRIDGE MODEL: SFC-45 - LOT# 1652251; FOAM TIP PLUNGER - LOT# 1643229; INJECTOR MODEL: MSI-PF - LOT# 1572654
Outcome Attributed to Adverse Event Required Intervention;
Patient Age21 YR
Patient SexMale
-
-