Model Number 776-LR4-HK-RP
Medical Device Problem Code Sharp Edges (4013)
Health Effect - Clinical Code Abrasion (1689)
Date of Event 10/20/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The carriere slx 3d clear brackets are ceramic, passive, self-ligating orthodontic brackets that directly bond to the teeth to provide a treatment solution for patients with malocclusions of primary, permanent, or mixed dentition.It was reported in the customer complaint form that bracket lr4 in the provided kit had sharp edges causing cheek irritation, cutting the cheek on the lower right of the patient and causing large sores that did not heal.The symptoms began the day after the bracket was placed.The bracket needed to be removed and replaced with a different brand.The 34-year old female patient was prescribed kenalog in orabase, a topical steroid.She was also advised to take otc pain medication such as ibuprofen and tylenol.The doctor indicated in the patient questionnaire that if the device malfunction were to recur, it would not likely cause or contribute to a death or serious injury.Also, no doctor or hospital visit was required, and as indicated in the questionnaire, the patient has fully recovered.Due to the fact that prescription and otc medications were given, ortho organizers decided to report this complaint to the fda.
 
Event or Problem Description
It was reported in the customer complaint form that bracket lr4 in the provided kit had sharp edges causing cheek irritation, cutting the cheek on the lower right of the patient and causing large sores that did not heal.The symptoms began the day after the bracket was placed.The bracket needed to be removed and replaced with a different brand.The 34-year old female patient was prescribed kenalog in orabase, a topical steroid.She was also advised to take otc pain medication such as ibuprofen and tylenol.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARRIERE SLX 3D CLEAR
Common Device Name
ORTHODONTIC CERAMIC BRACKET
Manufacturer (Section D)
ORTHO ORGANIZERS, INC.
1822 aston ave.
carlsbad CA 92008
Manufacturer (Section G)
ORTHO ORGANIZERS, INC.
1822 aston ave.
carlsbad CA 92008
Manufacturer Contact
nicole rajek
1822 aston ave.
carlsbad 92008
7604488600
MDR Report Key17847181
Report Number2081322-2023-00001
Device Sequence Number7247651
Product Code NJM
UDI-Device Identifier00190707057756
UDI-Public00190707057756
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K173440
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 09/29/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number776-LR4-HK-RP
Device Catalogue Number776-LR4-HK-RP
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/20/2023
Initial Report FDA Received Date09/29/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age34 YR
Patient SexFemale
-
-