| Catalog Number |
357.133 |
| Medical Device Problem Code |
Device-Device Incompatibility (2919)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
09/13/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: d9: complainant part is not expected to be returned for manufacturer review/investigation.H3, h4, h6: without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that a surgeon removed a tibial nail on (b)(6) 2023 because of a possible infection.The nail was difficult to remove for various reasons.The surgeon was unable to connect the appropriate extractor for the ex locking nails, and therefore had to try multiple options such as the non-synthes set, jamming two reaming rods, a non-synthes extraction hook, and a non-synthes conical extractor, which successfully extracted the nail.The non-synthes extraction hook tip broke off distally to the nail.There were already previous unknown fragments as well from when the surgeon was attempting to remove a difficult interlocking screw.The most distal screw was buried in bone and eventually was knocked out successfully.Additionally, one of the wrappers of a 3.0 reaming rod split.The surgery was completed successfully with a delay of 45 minutes.It is not known whether the fragments were removed from the tibia.After the surgery, it was noted in the sterile processing department that the extraction screw threads have been damaged.This report involves one extraction screw for ti femoral and tibial nails.This report is related to (b)(4), which reports the possible infection.This is report 2 of 2 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H6: device history record (dhr) review conducted: part number: 357.133.Lot number: 295p548.Manufacturing site: haegendorf.Release to warehouse date: 30th july 2021.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.G1: corrected physical manufacturer's name and address.H4: corrected device manufacture date.
|
| |
|
Search Alerts/Recalls
|
|