MAUDE Adverse Event Report: WELLSPECT HEALTHCARE NAVINA CLASSIC SYSTEM; NAVINA CLASSIC SYSTEM REGULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WELLSPECT HEALTHCARE NAVINA CLASSIC SYSTEM; NAVINA CLASSIC SYSTEM REGULAR Back to Search Results
Model Number 69005
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Abdominal Pain (1685); Hemorrhage/Blood Loss/Bleeding (1888); Sepsis (2067)
Date of Event 09/06/2023
Type of Reportable Event Serious Injury
Event or Problem Description
A female patient, had surgery and irradiation for colorectal cancer a few years ago.The patient has currently anal prolapse and has issues with frequency and incomplete bowel emptying, for which she was prescribed transanal irrigation.The patient got a product instruction on (b)(6) from a nurse trainer.A follow-up telephone call between the nurse and the patient followed on (b)(6).The patient said that she was bleeding.The nurse advised her to have this urgently clarified by a doctor.The incident reported was on (b)(6), the nurse does not know whether the patient followed the advice and saw a doctor or continued the therapy between on (b)(6) since she could not reach the patient by phone during this week.The patient was alone when using the device and the rectal catheter balloon did not burst inside the rectum which normally is a cause of bowel perforation.According to the reported complaint, the patient injured her bowel so badly that emergency surgery was required.A large piece of the intestine was removed.The patient now has an artificial outlet.She was in an artificial coma (sedated), and she has been brought out of the coma.Because intestinal contents leaked into the abdominal cavity, there was severe sepsis plus damage to organs such as the heart, kidneys, and liver.According to the patient's daughter, the anastomosis from the colorectal cancer surgery is not the reason for the perforation.
 
Additional Manufacturer Narrative
The product and product lo.No.Was not available to be returned.Without the benefit of examination and testing, wellspect healthcare is precluded from commenting on the condition of the device or the cause of the occurrence.Should additional facts prompt us to alter or supplement any information of conclusions contained in this report, a follow-up report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NAVINA CLASSIC SYSTEM
Common Device Name
NAVINA CLASSIC SYSTEM REGULAR
Manufacturer (Section D)
WELLSPECT HEALTHCARE
aminogatan 1
p.o.box 14
molndal, SE-43 121
SW  SE-43121
Manufacturer (Section G)
WELLSPECT HEALTHCARE
aminogatan 1
p.o.box 14
molndal, SE-43 121
SW   SE-43121
Manufacturer Contact
emad ramzi
aminogatan 1
p.o.box 14
molndal SE-43-121
SW   SE-43121
MDR Report Key17849685
Report Number3009632672-2023-00006
Device Sequence Number20522883
Product Code KNT
UDI-Device Identifier07392532198434
UDI-Public07392532198434
Combination Product (Y/N)N
Initial Reporter CountrySW
PMA/510(K) Number
K190977
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Consumer
Initial Reporter Occupation Other
Remedial Action Patient Monitoring
Type of Report Initial
Report Date (Section B) 10/02/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model Number69005
Device Catalogue Number6900515
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/11/2023
Initial Report FDA Received Date10/02/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
Patient Age79 YR
Patient SexFemale
Patient RaceWhite
-
-