SMITH & NEPHEW, INC. DELTA HEAD 36MM 12/14 SHORT +0MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER, CEMENTED OR NON-POROUS Back to Search Results
Catalog Number 76539165
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Medical device site infection (4845)
Date of Event 09/12/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that, after thr surgery had been performed on an unknown date, the patient experienced an infection.This adverse event was solved by an intervention on (b)(6) 2023, in which a dair (debridement, antibiotics and implant retention) procedure was performed with the exchange of the r3 36mm id intl dlt cer lnr 54mm and the delta head 36mm 12/14 short +0mm.Patient's health status is unknown.No further complications were reported.
 
Additional Manufacturer Narrative
H10: internal complaint reference: (b)(4).
 
Additional Manufacturer Narrative
D11: concomitant medical products and therapy dates.The devices were not returned for evaluation; therefore, a device analysis could not be performed.The clinical/medical investigation concluded that, as of the date of this medical investigation, the requested clinical documentation has not been provided; therefore, there were no clinical factors found which would have definitively contributed to the event.The patient impact beyond the reported infection and additional procedure cannot be determined.The patient¿s current health status was not provided.Therefore, no further medical assessment can be rendered.A review of the production orders did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history for the part numbers over the past 12 months and for the batch numbers based on historical data of the devices did not reveal similar events for the listed device.A review of the instructions for use documents for total hip systems revealed that infection, both early, post-operative superficial and early, post-operative deep wound infection and late periprosthetic infection has been identified in potential complications associated with total hip arthroplasty surgery, primary or revision section.A review of the risk management files revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to these products and event.A review of the sterilization records revealed the batch was sterilized within normal parameters.At this time, we have no reason to suspect that the products failed to meet any specifications at the time of manufacture.Factors that could contribute to the reported event include loss of sterility during procedure, injury, patient condition and/or medical history.Based on this investigation, the need for corrective action is not indicated.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DELTA HEAD 36MM 12/14 SHORT +0MM
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER, CEMENTED OR NON-POROUS
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key17851679
Report Number1020279-2023-01864
Device Sequence Number6759204
Product Code LZO
UDI-Device Identifier03596010566959
UDI-Public03596010566959
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K083762
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number76539165
Device Lot Number16LT77554
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/12/2023
Supplement Date Received by Manufacturer11/02/2023
Initial Report FDA Received Date10/02/2023
Supplement Report FDA Received Date11/04/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/02/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
R3 36MM ID INTL DLT CER LNR 54MM, LOT#:16HT74092
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-