EXACTECH, INC. EQUINOXE, HUMERAL STEM PRIMARY, PRESS FIT 13MM; PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED
|
Back to Search Results |
|
| Model Number |
RS GLENOID PLATE EXT CAG +10MM CAGE PEG |
| Medical Device Problem Codes |
Fracture (1260); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Bone Fracture(s) (1870); Periprosthetic fracture (4902)
|
| Date of Event |
09/14/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
As reported, the patient had an initial right tsa on (b)(6) 2023.The patient presented - in taking off the humeral head, the stem became loose and the portion proximally which had stress shielding broke free from shaft.The case report form indicates that this event is definitely not related to device, definitely related to procedure.Outcome: resolved on (b)(6) 2023.
|
| |
|
Additional Manufacturer Narrative
|
|
Pending investigation.D10: serial number.Item number and full description.(b)(6), 320-38-00 - 145-deg pe 38mm hum liner +0.(b)(6), 320-06-38 - glenosphere 38mm.(b)(6), 304-22-11 - 10.5mm platform fx stem right.(b)(6), 320-10-00 - equinoxe reverse tray adapter plate tray +0.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional and/or corrected information.The following sections were updated/corrected: b3, d1, d4, g4, h4, h6.Mdr section codes updated/corrected: a, b, c, d, e, f, g.Implant date is unknown.The reason for the intraoperative bone fracture reported cannot be conclusively determined but may be related to a patient condition or high forces during broaching, reaming, exposure, or retraction.However, this cannot be confirmed based on the information provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|