|
DEPUY SPINE INC POWERED INSTRUMENTS EXPEDIUM SYSTEM T-HANDLE, RATCHETING, QUICK COUPLING 5.5; INSTRUMENT, MANUAL, SURGICAL, GENERAL USE
|
Back to Search Results |
|
| Catalog Number |
279703150P |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Device report from canada reports an event as follows: it was reported that during an unknown procedure on an unknown date in 2023, the handle in question broke during use.No fragments were generated, there was no reported adverse patient consequence.The procedure was completed successfully with no surgical delay.No additional information is available in report.This report is for a powered instruments expedium system t-handle, ratcheting, quick coupling 5.5.This is report 1 of 1 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.B3: event occurred on an unknown date in 2023.D9: complainant part is not expected to be returned for manufacturer review/investigation.H3, h4, h6: without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|
|