| Catalog Number |
03666794001 |
| Medical Device Problem Codes |
Image Display Error/Artifact (1304); No Device Output (1435)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/05/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The test strips' lot number was 71519013 with an expiration date of 31-dec-2024.During the troubleshooting, a display check was performed and revealed the following: 888 appeared clear and complete.The hourglass was faded.The meter memory result appeared faded on the sides.The investigation is ongoing.
|
| |
|
Event or Problem Description
|
|
We received an allegation about a coaguchek xs meter having a faded display and not detecting the test strips.The reporter stated that they had to tilt the meter in order to see everything on the display because it was faded.The reporter also confirmed that no harm occurred in relation to the display issues because the meter had not been in use since the issues began.
|
| |
|
Additional Manufacturer Narrative
|
|
The meter was not received for investigation.
|
| |
|