| Catalog Number |
4290 |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Loose or Intermittent Connection (1371); Air/Gas in Device (4062)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the needle was too loose on the safety system, which meant that air enters the syringe when it is drawn up and that fluid leaks next to the needle when injecting.The needle spontaneously comes loose from the safety system during the puncture.Adverse patient effects are unknown.It was reported that no further information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
Other text: b3: date of event is unknown; no information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation codes: updated.D3, g1,2 email is: regulatory.Responses@icumed.Com.No product sample has been returned.The customer provided two photos, one identifying the lot number and the second showing the product configuration.Visual inspection of the pictures did not reveal any defects or anomalies.Without a product sample the reported problem could not be verified and/or confirmed with confidence; therefore, no further action will be conducted at this time.It is possible that the issues observed by the customer may have occurred if the mating luers of the components were not seated as instructed within the instruction for use (ifu).A device history record (dhr) review reported no discrepancies or non-conformances during the manufacturing of the reported lot number.Complaint information will continue to be monitored for any new information or adverse trends and further actions will be taken accordingly.
|
| |
|
Search Alerts/Recalls
|