| Model Number |
500-55112 |
| Medical Device Problem Code |
Obstruction of Flow (2423)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/03/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the lytic therapy would not infuse.During a bilateral ekos case, two ekos+ devices were inserted into the patient.When the ekos+ kit 106cm 12 cm was hooked up to infuse lytic, the drug would not infuse.The drug line was aspirated in an attempt to troubleshoot the issue.The ekos+ device was replaced with an ekosonic endovascular device, 106x12cm to troubleshoot the issue, but the new ekosonic device would not infuse lytic.Once again, the drug line was aspirated in an attempt to troubleshoot the issue.As a result of the device malfunction, the physician elected to remove the ekosonic endovascular device, and the patient was transferred to the icu with only one ekos+ device in place.It was unknown whether the patient received the correct dose of lytic therapy.However, there was no reported undesirable patient outcome.
|
| |
|
Additional Manufacturer Narrative
|
|
Device analysis: the drug lumen was hooked up to a non-boston scientific infusion pump and ran for a sustained 30 minutes of infusion.No pump error alarms were noticed.A syringe filled with blue dye was attached to the drug luer and flushed through the catheter under water.The blue dye was able to be seen coming from the drug holes and confirmed the ability of the device to be flushed.The reported event of obstruction within the drug line was unable to be confirmed.
|
| |
|
Event or Problem Description
|
|
It was reported that the lytic therapy would not infuse.During a bilateral ekos case, two ekos+ devices were inserted into the patient.When the ekos+ kit 106cm 12 cm was hooked up to infuse lytic, the drug would not infuse.The drug line was aspirated in an attempt to troubleshoot the issue.The ekos+ device was replaced with an ekosonic endovascular device, 106x12cm to troubleshoot the issue, but the new ekosonic device would not infuse lytic.Once again, the drug line was aspirated in an attempt to troubleshoot the issue.As a result of the device malfunction, the physician elected to remove the ekosonic endovascular device, and the patient was transferred to the icu with only one ekos+ device in place.It was unknown whether the patient received the correct dose of lytic therapy.However, there was no reported undesirable patient outcome.
|
| |
|
Search Alerts/Recalls
|