| Model Number |
510A1960NF |
| Medical Device Problem Code |
Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
B3: date of event and d4: udi number is unknown; no information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported that there was a malfunction during the transfer of a newborns.Adverse patient effects are unknown and the outcome of the event was listed as resolved.
|
| |
|
Additional Manufacturer Narrative
|
|
Other, other text: h6: evaluation codes updated.Device evaluation: one device was received.A visual inspection found the device in good condition except for the damaged patient valve.During the functional testing it was found that the alarm was not working reliably and there was a damaged patient valve that caused peep limits out of tolerance.Also the alarm reed is not squeaking.The electronics will be checked or replaced, the damaged patient valve will be replaced, and the alarm reed will be checked or replaced.Service history review identified there was no indication that the complaint was related to a service of the device within the review period.For all enquiries or follow-up questions related to the record, do not use regulatory.Responses@smiths-medical.Com located in sections g.1., please direct those to the following: regulatory.Responses@icumed.Com.
|
| |
|
Additional Manufacturer Narrative
|
|
Udi related data quality updates only.
|
| |
|