• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OSTEOMED, LLC LEVER ARM MOTOR UNIT; HANDPIECE, ROTARY BONE CUTTING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OSTEOMED, LLC LEVER ARM MOTOR UNIT; HANDPIECE, ROTARY BONE CUTTING Back to Search Results
Model Number 450-0034
Medical Device Problem Code Power Problem (3010)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
Service request for the device was received with no product complaint made.On (b)(6) 2023.The service request work order device diagnosis was completed.During the service request work order, it was noted the device was running continuously due to a malfunctioning cable assembly, and motor unit also has worn and dirty rotor assembly".As a result of these findings of the service request work order, this report is being submitted.Additionally, as a result of the service request work order, the customer was contacted to determine if any surgery or patient was impacted.The customer responded that this issue was noted in their warehouse post surgery with no patient, user, or surgery involved.
 
Additional Manufacturer Narrative
The device had been returned under a service work order (sro) during which it was identified that the device was continuously running due to malfunction of the cable assembly.The device is a reusable instrument and was manufactured in 2020; therefore, it is unknown how many times the device was used prior to this reported complaint event nor under what circumstances it was used.The device was tested per otr 0060 rev e upon manufacture and met all specifications.Based on the information received and the investigation performed, the root cause could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVER ARM MOTOR UNIT
Common Device Name
HANDPIECE, ROTARY BONE CUTTING
Manufacturer (Section D)
OSTEOMED, LLC
3885 arapaho rd
addison TX 75001
Manufacturer (Section G)
OSTEOMED, LLC
3885 arapaho rd
addison TX 75001
Manufacturer Contact
ellie wood
3885 arapaho rd
addison, TX 75001
9726774600
MDR Report Key17857713
Report Number2027754-2023-00034
Device Sequence Number17676059
Product Code KMW
UDI-Device Identifier00845694015646
UDI-Public(01)00845694015646(30)1(11)200127
Combination Product (Y/N)N
Initial Reporter CountryID
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 10/02/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number450-0034
Device Catalogue Number450-0034
Device Lot Number1145682
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/07/2023
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/07/2023
Initial Report FDA Received Date10/02/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/27/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-