| Model Number |
MC162167S |
| Medical Device Problem Codes |
Break (1069); Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/10/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the physician was doing a face arteriovenous malformation (avm case) using a liquid embolic.As the physician removed the microcatheter after approximately 15 minutes of use, he realized that the outer tip of the microcatheter was missing.Using an ultrasound post procedure, as the tip was not visible on the angiogram, the physician confirmed that the tip of the microcatheter was found hanging on the wall of the carotid.No intervention was done post procedure, and the tip was left inside the patient.There was no reported injury to the patient and was reported to be good.
|
| |
|
Additional Manufacturer Narrative
|
|
A search for non-conformances associated with this part / lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.The device was reported to be available for return to the manufacturer for evaluation but has not yet been returned.The reported issue could not be confirmed.If the device is received at a later date, an investigation will be performed and a supplemental mdr will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported complaint is confirmed as the condition of the returned device is consistent with the reported complaint.The cause of the broken microcatheter could not be determined from the investigation but is consistent with experiencing forces over specification during the procedure.
|
| |
|
Event or Problem Description
|
|
See h10.
|
| |
|