ROCHE DIAGNOSTICS ELECSYS FERRITIN; FERRITIN ASSAY Back to Search Results
Catalog Number 03737551160
Medical Device Problem Codes Low Test Results (2458); Non Reproducible Results (4029)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/30/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The serial number of the customer's cobas e411 rack is (b)(6).The customer stated that the pinch valve tubes were in very bad condition and had not been replaced for more than six months.The customer replaced the pinch valve tubes after a repeat result was obtained.The investigation is ongoing.
 
Event or Problem Description
The initial reporter received a questionable result elecsys ferritin (ferritin) from one patient sample tested on the cobas e411 rack.The initial result was not reported outside the laboratory.A rerun was requested by the laboratory results validation department due to an initial result not matching the patient history.The patient's other sample was also run.On (b)(6) 2023: the initial result of the initial sample was 0.638 ng/ml with a data flag.The result of the other sample was 331.8 ng/ml.On (b)(6) 2023: the repeat result of the initial sample was 336.4 ng/ml.The initial result was deemed an incorrect result.
 
Additional Manufacturer Narrative
The investigation determined the event was caused by the worn pinch valve tubes.After service, no further issues were reported by the customer.The investigation determined the service actions resolved the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS FERRITIN
Common Device Name
FERRITIN ASSAY
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhofer strasse 116
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
indianapolis, IN 46250
MDR Report Key17859481
Report Number1823260-2023-03158
Device Sequence Number5730211
Product Code JMG
UDI-Device Identifier04015630938940
UDI-Public04015630938940
Combination Product (Y/N)Y
Initial Reporter CountryEG
PMA/510(K) Number
K971833
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2023
Device Catalogue Number03737551160
Device Lot Number64725205
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 09/07/2023
Supplement Date Received by Manufacturer10/23/2023
Initial Report FDA Received Date10/03/2023
Supplement Report FDA Received Date11/13/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-