| Model Number |
MI1200 SYNCHRONY |
| Medical Device Problem Code |
Impedance Problem (2950)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The user's hearing performance with the device is affected after a trauma to the left side of the head.The user's parents also reported redness at the side of the implant.Reimplantation is considered.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has not been explanted.If it should be explanted, it should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: based on the received information, a damage to the active electrode due to reported mechanical impact appears very likely.In addition in situ measurements before the reported impact already show one channel with hi status due to undetermined reasons.However, to determine an exact root cause device investigation would be necessary.Re-implantation is considered but no date has been scheduled yet.
|
| |
|
Event or Problem Description
|
|
The user's hearing performance with the device is affected after a trauma to the left side of the head.The user's parents also reported redness at the side of the implant.Reimplantation is considered.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusion: overload fractures in the active electrode which are consistent with an external mechanical impact were determined to be the root cause of device failure.The active electrode also showed some additional damage to the electrode wires, as being caused by minute device mobility.An accident or impact might have weakened the fixation of the implant, consequently leading to electrode mobility.The problems given in the recipient report appear to match the damage found.This is a final report.
|
| |
|
Event or Problem Description
|
|
The user's hearing performance with the device is affected after a trauma to the left side of the head.The user's parents also reported redness at the side of the implant.The user has been re-implanted.
|
| |
|