MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 SENSICA UO SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 SENSICA UO SYSTEM Back to Search Results
Catalog Number SCCS1002E
Medical Device Problem Codes Defective Component (2292); Failure to Calibrate (2440)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/12/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
The sensica urine output system touchscreen need update, and failed the during screen calibration.
 
Additional Manufacturer Narrative
The reported issue was confirmed.The root cause identified is a failed tablet/pc.The device was evaluated upon receipt.The sensica touchscreen failed the during calibration, and testing.Replaced tablet/pc.Other maintenance performed.Applied the optical sensor gasket foam, the tamper evident labels, and new ground wire.Completed final functional.Dhr is not required as the reported event is not an out of box failure and therefore the reported event is not manufacturing related.The labelling review sis not required as the complaint or reported issue was confirmed through other elements of the investigation to not be labeling or packaging related.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the actual/suspected device was inspected.
 
Event or Problem Description
The sensica urine output system touchscreen need update, and failed the during screen calibration.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENSICA UO SYSTEM
Common Device Name
SENSICA UO SYSTEM
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key17860088
Report Number1018233-2023-07147
Device Sequence Number20379167
Product Code EXS
UDI-Device Identifier00801741219016
UDI-Public00801741219016
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberSCCS1002E
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/12/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/12/2023
Supplement Date Received by Manufacturer03/15/2024
Initial Report FDA Received Date10/03/2023
Supplement Report FDA Received Date03/18/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/01/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-