| Model Number |
H403023 |
| Medical Device Problem Code |
Output Problem (3005)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/14/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
During a supraventricular tachycardia procedure, the stimulator touchscreen showed an "ep-4 is off" error and switched back to the menu screen instead of a screen to pace from.The issue was temporarily resolved by restarting the recording system, stimulator and touchscreen, however the issue recurred.The stimulator was replaced, and the issue was resolved.There were no adverse consequences to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.No non-conformances associated with the reported event were identified.No capa is required; the root cause of the reported issue could not be confirmed to be a manufacturing, design, or quality system related occurrence.The reported issue will continue to be monitored for trends and reviewed in qdr (quality data review).Based on the information received, the cause of the reported output issue and subsequent delay remains unknown.
|
| |
|