| Catalog Number |
08057800190 |
| Medical Device Problem Code |
Low Test Results (2458)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/05/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
There was an allegation of questionable gluc3 glucose hk gen.3 results for 1 patient sample on a cobas pro c 503 analytical unit.The initial glucose result was 21.7 mg/dl.The customer questioned the low result and repeated the sample.The repeat result was 83.7 mg/dl.No questionable results were reported outside of the laboratory.
|
| |
|
Additional Manufacturer Narrative
|
|
The analyzer serial number is (b)(6).The field service engineer (fse) changed the sample probe, replaced the heat cut filter, adjusted the rinse levels, cleaned the reagent probes and a solenoid valve, and performed a cell blank successfully.The customer performed qc successfully.The investigation is ongoing.
|
| |
|
Additional Manufacturer Narrative
|
|
The service maintenance actions performed by the fse resolved the issue.No further issues were reported after the service visit.
|
| |
|