| Model Number |
N051 |
| Medical Device Problem Code |
Defective Device (2588)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/10/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) recorded a red call doctor alert via the associated remote communicator, indicative a device condition requiring urgent attention.Technical services (ts) recommended the patient contact their clinic for review.This device remains in service.No adverse patient effects were reported.
|
| |
|
Event or Problem Description
|
|
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) recorded a red call doctor alert via the associated remote communicator, indicative a device condition requiring urgent attention.Technical services (ts) recommended the patient contact their clinic for review.This device was subsequently explanted and successfully replaced.No adverse patient effects were reported.
|
| |
|
Event or Problem Description
|
|
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) recorded a red call doctor alert via the associated remote communicator, indicative a device condition requiring urgent attention.Technical services (ts) recommended the patient contact their clinic for review.This device was subsequently explanted and successfully replaced.No adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The returned device was thoroughly inspected and analyzed upon receipt at our post market quality assurance laboratory.The device was exposed to simulated heart load conditions, and the defibrillation pacing, and sensing functions were tested.Impedance testing was completed and all measurements were within normal limits.The device operated appropriately with no interruptions in therapy output at the returned programmed settings.A series of electrical tests was also performed, and again, normal device function was observed.Analysis did not identify any device characteristics that would have caused or contributed to the reported clinical observations.
|
| |
|