| Model Number |
VTICM5_12.1 |
| Medical Device Problem Codes |
Off-Label Use (1494); Inadequacy of Device Shape and/or Size (1583); Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/28/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The reporter indicated that the surgeon implanted a 12.1mm vticm5_12.1, -11.0/1.5/179 (sphere/cylinder/axis), implantable collamer lens into the patient's right eye (od) on (b)(6)2023.Lens was repositioned on (b)(6) 2023 but this did not resolve the lens rotation and low vault.The lens was replaced with a longer length spherical lens on (b)(6) 2023.This did not resolve the problem.Cause of the event is reported as patient related factor, lens too short cause rotation.Reportedly, patient states near vision not so clear.
|
| |
|
Additional Manufacturer Narrative
|
|
A4 - unk.A5 - unk.A6 - unk.H6 - work order search: no similar complaint type events were reported for units within the same lot.Claim# (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
H6: device code: 1494: this lens model is contraindicated for patients with an anterior chamber depth (acd) less than 3.0mm.Claim#: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
H3: device evaluation lens was returned in vial in liquid.Visual inspection found no visible damage to lens.Dimensional inspection found the lens to be within specification.(b)(4).
|
| |
|