• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC INTRAOCULAR LENS Back to Search Results
Model Number VICM5 13.2
Medical Device Problem Codes Inadequacy of Device Shape and/or Size (1583); Patient Device Interaction Problem (4001)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/08/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Claim# (b)(4).
 
Event or Problem Description
The reporter indicated that the surgeon implanted a 13.2mm vicm5_13.2, -10.5 diopter, implantable collamer lens into the patient's right eye (od) on (b)(6) 2023.The lens was replaced with a longer length lens due to low vault on (b)(6) 2023.This resolved the problem.Cause of the event is reported as patient related factor, lens too short.Reportedly, vault and near vision improved.
 
Additional Manufacturer Narrative
B5 - the problem was resolved.Reportedly, " better vault after lens exchanged.To monitor vault and patient's vision quality.Patient has no further complaint." claim# (b)(4).
 
Additional Manufacturer Narrative
H3: device evaluation lens was returned in vial in liquid.Visual inspection found no visible damage to lens.Dimensional inspection found the lens to be within specification.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Common Device Name
PHAKIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
joselene muniz
1911 walker avenue
monrovia, CA 91016
MDR Report Key17862045
Report Number2023826-2023-04224
Device Sequence Number20259967
Product Code MTA
UDI-Device Identifier00840311300945
UDI-Public00840311300945
Combination Product (Y/N)N
Initial Reporter CountrySN
PMA/510(K) Number
P030016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/28/2026
Device Model NumberVICM5 13.2
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/15/2023
Supplement Date Received by Manufacturer08/15/2024
10/02/2024
Initial Report FDA Received Date10/03/2023
Supplement Report FDA Received Date09/11/2024
10/24/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/17/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNK.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age30 YR
Patient SexFemale
-
-