| Model Number |
VICM5 13.2 |
| Medical Device Problem Codes |
Inadequacy of Device Shape and/or Size (1583); Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/08/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Claim# (b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated that the surgeon implanted a 13.2mm vicm5_13.2, -10.5 diopter, implantable collamer lens into the patient's right eye (od) on (b)(6) 2023.The lens was replaced with a longer length lens due to low vault on (b)(6) 2023.This resolved the problem.Cause of the event is reported as patient related factor, lens too short.Reportedly, vault and near vision improved.
|
| |
|
Additional Manufacturer Narrative
|
|
B5 - the problem was resolved.Reportedly, " better vault after lens exchanged.To monitor vault and patient's vision quality.Patient has no further complaint." claim# (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
H3: device evaluation lens was returned in vial in liquid.Visual inspection found no visible damage to lens.Dimensional inspection found the lens to be within specification.(b)(4).
|
| |
|