• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NUVASIVE, INCORPORATED NUVASIVE GENERAL INSTRUMENTS; ORTHOPEDIC MANUAL SURGICAL INSTRUMENTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NUVASIVE, INCORPORATED NUVASIVE GENERAL INSTRUMENTS; ORTHOPEDIC MANUAL SURGICAL INSTRUMENTS Back to Search Results
Model Number 6940429
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/20/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was returned and the complaint was confirmed.A review of the manufacturing history found no material nonconformance, no manufacturing error, nor discrepancies with respect to material type, treatments or dimension that may have caused or contributed to this event.Parts met acceptance criteria upon release.Lot #fd9127220 was manufactured in october of 2019.Examination of the returned device was completed by nuvasive and the fracture pattern identified is consistent with excessive force.Review of the returned device and reported event suggests the fracture was the result of off-angled excessive force during use and considered physical field damage.No additional investigation necessary.Labeling review: ".Warnings, cautions and precautions: the implantation of spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient." ".Do not implant the instruments: complications to the patient may include, but are not limited to: breakage of the device, which could make necessary removal difficult or sometimes impossible, with possible consequences of late infection and migration.Breakage could cause injury to the patient." ".Pre-operative warnings: the method of use for the instruments are to be determined by the user¿s experience and training in surgical procedures.The instruments should be treated as any precision instrument and should be carefully placed on trays, cleaned after each use, and stored, according to generally accepted hospital methods and practices.The instruments should be carefully examined prior to use for functionality, excessive wear, or damage.A damaged instrument should not be used as this may increase the risk of malfunction and potential patient injury.Care should be used during surgical procedures to prevent damage to the devices and injury to the patient." ".Intra-operative warnings: the physician should take precautions against putting undue stress on the spinal area with instruments.Any surgical technique should be carefully followed.It is important that the surgeon exercise extreme caution when working in close proximity to vital organs, nerves, or vessels, and that the force applied to the instrumentation is not excessive, to prevent potential injury to the patient.Over-bending, notching, striking, and/or scratching of implants with any instrument should be avoided to reduce the risk of breakage.The physical characteristics required for many instruments do not permit them to be manufactured from implantable materials.If any broken fragments of instruments remain in the body of a patient, they could cause allergic reactions or infections.If an instrument breaks in surgery and fragments go into the patient, these pieces should be removed prior to closure and should not be implanted.".
 
Event or Problem Description
It was reported that during an anterolateral posterior fusion procedure at l4/5, distal tip of the pituitary broke off.The fractured portion was not left in the patient; the surgeon was able to retrieve the fractured piece from the patient's body using an extra-long xlif suction tip.There was no adverse impact to the patient.No additional information is available.
 
Event or Problem Description
N/a.
 
Additional Manufacturer Narrative
Correction: d2.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NUVASIVE GENERAL INSTRUMENTS
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENTS
Manufacturer (Section D)
NUVASIVE, INCORPORATED
7475 lusk blvd
san diego CA 92121
Manufacturer (Section G)
NUVASIVE, INCOPORATED
7475 lusk blvd
san diego CA 92121
Manufacturer Contact
geoffrey gannon
7475 lusk blvd
san diego, CA 92121
MDR Report Key17863035
Report Number2031966-2023-00218
Device Sequence Number6135454
Product Code LXH
UDI-Device Identifier00887517452764
UDI-Public887517452764
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
NA
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/15/2024
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number6940429
Device Lot NumberFD4809323
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/20/2023
Supplement Date Received by Manufacturer02/15/2024
Initial Report FDA Received Date10/03/2023
Supplement Report FDA Received Date02/15/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/31/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexPrefer Not To Disclose
-
-