| Model Number |
OSTETOME, ANGLED UP, 12MM |
| Medical Device Problem Code |
Defective Device (2588)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/05/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On 9/5/2023, it was reported by a sales representative via sems-06022369 that an ar-8970-13 osteotome, angled up 12 mm was damaged.The item came from a set and was discovered during a case.There were no adverse effects to the patient.This occurred during use in an unspecified procedure with no patient harm.Additional information has been requested.
|
| |
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Complaint confirmed.One unpackaged ar-8970-13, batch 041439 was received for investigation.The visual evaluation revealed that the received device's blade tip had nicks and was damaged.The most likely cause(s) of this type of event include applying excessive forces through leveraging/prying the device.
|
| |
|