| Model Number |
2.7MM SCREW DEPTH INDICATOR |
| Medical Device Problem Code |
Fitting Problem (2183)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/05/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the length measurement did not work properly and therefore wrong screws were chosen.There was no harm for patient, operator or third party reported.No further information received.***update 13-sep-2023 nroe: further information revealed that this incident occurred during a clavicle fracture treatment with a distal clavicle plate.There was no harm for patient, operator or third party.The surgery was finished successfully after replacing the screws that were to long with shorter screws.It was not necessary to switch the surgical technique or do a second surgery.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d9, g3, h3, h6.The complaint is confirmed.One unpackaged ar-2670 batch 1068642103 was received for evaluation.Upon visual evaluation, the measuring needle tip was bent.Wear and tear were also observed.Measurement of the instrument's overall length by drawing c11016 at revision 3 found the device out of tolerance due to the bent on the tip.No functional test was performed due to the damage to the device.This event is most likely caused by misuse by prying/levering the device against hard bone or other instrumentation during use.
|
| |
|
Search Alerts/Recalls
|