• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. SFX SYM PDS+ UNI VIO 18IN 1 S/A CTXB; SUTURE, SURGICAL, ABSORBABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. SFX SYM PDS+ UNI VIO 18IN 1 S/A CTXB; SUTURE, SURGICAL, ABSORBABLE Back to Search Results
Catalog Number SXPP1A300
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).H6 component code: g07002 - device not returned.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.The following information was requested, but unavailable: please provide the lot number.=unk.Event/procedure name and date?=unk.Please provide the source or name and title of external person providing answers to follow-up (external person submitting answers to sales rep)= (b)(4).
 
Event or Problem Description
It was reported that a patient underwent an unknown procedure on an unknown date and barbed suture was used.During the procedure, when putting the 1st stitch, the fixation tab was broken.Further details are not provided from the hp.When we send the sample to you, we will let you know its return date and tracking number.No adverse patient consequences were reported.Additional information was requested.
 
Additional Manufacturer Narrative
Product complaint#: (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H3: investigation summary: the product was returned for evaluation.Visual inspection was conducted on the returned device.The returned sample determined that it was received, one open sample that pertained to the product code sxpp1a300.During the visual assessment of the needle suture, it was noted that the swage and attachment area was as expected.Marks on the body needle that appears to be caused by a surgical instrument could be observed.The suture was examined, body fluids at some barbs were found and the sides of the fixation tabs were found to be broken.It should be noted that a 100% inspection is performed via the automotive vision system to ensure the tab is present and complete during manufacturing.As part of the quality process, all devices are manufactured, inspected, and released to approved specifications.The instructions for use contain the following caution: as with any device, care should be taken to avoid damage to the strand when handling.Avoid the crushing or crimping action of the surgical instruments, such as needle holders and forceps.To seat the fixation tab; pull the device through the tissue to gently seat the fixation tab against the tissue.The fixation tab should be seated above the tissue plane and be visible.Do not exert additional force on the fixation tab.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SFX SYM PDS+ UNI VIO 18IN 1 S/A CTXB
Common Device Name
SUTURE, SURGICAL, ABSORBABLE
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-JUAREZ
calle durango #2751
ciudad juarez 32575
MX   32575
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
9083863534
MDR Report Key17865648
Report Number2210968-2023-07375
Device Sequence Number15214258
Product Code NEW
UDI-Device Identifier10705031219359
UDI-Public10705031219359
Combination Product (Y/N)Y
Initial Reporter CountryJA
PMA/510(K) Number
K113004
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberSXPP1A300
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/07/2023
Supplement Date Received by Manufacturer10/18/2023
Initial Report FDA Received Date10/03/2023
Supplement Report FDA Received Date11/10/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-