Catalog Number 502-03-56E
Medical Device Problem Codes Loss of Osseointegration (2408); Osseointegration Problem (3003)
Health Effect - Clinical Codes Pain (1994); Inadequate Osseointegration (2646)
Date of Event 09/11/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.H3 other text : not available.
 
Event or Problem Description
It was reported that the patient's hip was revised.As reported: "patient was revised due to pain and suspected acetabular component loosening.During surgery it was noted that the acetabular component was grossly loose.There was some presence of corrosive changes on the trunnion as well as within the femoral head.Surgeon elected to retain the femoral component and place a new acetabular component.A sleeved ceramic head was implanted.".
 
Additional Manufacturer Narrative
Reported event: an event regarding loosening involving a trident shell was reported.The event was confirmed via clinician review of the provided medical records.Method & results: product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.Clinician review: a review of the provided medical information by a clinical consultant indicated: "the ap pelvis x-ray shows that the right tha acetabular component has spun in to a vertical position confirming lack of fixation.No documentation was available to confirm that a revision was completed.The root cause of the acetabular loosening cannot be determined from the limited documentation provided." product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: it was reported that the patient was revised due to loosening of the shell.Intraoperatively, some presence of corrosive changes on the trunnion as well as within the femoral head were observed.The femoral head was revised in addition to the acetabular component.The shell loosening event was confirmed via clinician review of the provided medical records.The exact cause of the event could not be determined because insufficient information was provided.Further information such as pathology reports, additional pre- and post-operative x-rays as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If devices and/or additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Event or Problem Description
It was reported that the patient's hip was revised.As reported: "patient was revised due to pain and suspected acetabular component loosening.During surgery it was noted that the acetabular component was grossly loose.There was some presence of corrosive changes on the trunnion as well as within the femoral head.Surgeon elected to retain the femoral component and place a new acetabular component.A sleeved ceramic head was implanted.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRIMARY TRITANIUM HEMI CLUSTER HOLE CUP 56MM
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
arokiya raj
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key17866031
Report Number0002249697-2023-01111
Device Sequence Number5493601
Product Code LPH
UDI-Device Identifier07613327040166
UDI-Public07613327040166
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K143085
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2012
Device Explanted Year2023
Reporter Type Manufacturer
Report Source Other,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/31/2017
Device Catalogue Number502-03-56E
Device Lot NumberMLJVNY
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 09/11/2023
Supplement Date Received by Manufacturer11/08/2023
Initial Report FDA Received Date10/03/2023
Supplement Report FDA Received Date11/30/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/21/2012
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age77 YR
Patient SexFemale
Patient Weight69 KG
-
-