| Catalog Number |
305125 |
| Medical Device Problem Code |
Packaging Problem (3007)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/19/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
B.3.Date of event: unknown.The date received by manufacturer has been used for this field.H.3.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported bd® needle 1 in.Single use, sterile, had mixed product in the package.The following was received by the initial reporter: the issue that one of their sites reported that there are 14 0.5 inch needles that were packed with the 1 inch needles which were separated from the 1 inch needles.
|
| |
|
Additional Manufacturer Narrative
|
|
H.6.Investigation summary: a device history record review was completed by our quality engineer team for provided material number 305125 and lot number 2117120.The review did not reveal any detected abnormalities during the production process that could have contributed to this defect and all quality tests were found to be within specification.As a sample was unavailable for return, a thorough sample investigation could not be completed.Based on the investigation results, an exact cause for this incident could not be identified.There are quality controls currently in place to detect this type of defect during the production process.Further action has not been determined necessary at this time.Complaints received for this device and reported condition will continue to be tracked and trended.Our quality team regularly reviews the collected data for identification of emerging trends.H3 other text : see h10.
|
| |
|
Event or Problem Description
|
|
It was reported bd® needle 1 in.Single use, sterile, had mixed product in the package.The following was received by the initial reporter: the issue that one of their sites reported that there are 14 0.5 inch needles that were packed with the 1 inch needles which were separated from the 1 inch needles.
|
| |
|