Catalog Number 5990011
Medical Device Problem Code Improper or Incorrect Procedure or Method (2017)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 03/28/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
As reported, the echo 2 ps portion of the ventralight st w/ echo 2 ps was inadvertently left in the patient following mesh implant; the patient was taken back to the or and the portion was removed.No additional complications were reported.Based on the information provided the reported event is determined to be use related with no malfunction of the device as the instructions-for-use were not followed.Per the instructions-for-use, supplied with the device, ¿ventralight¿ st mesh is the only permanent implant component of the device.The echo 2¿ positioning system (which includes deployment frame, center hoisting suture and all connectors) must be removed from the patient and appropriately discarded.It is not part of the permanent implant.¿ "verify that the echo 2¿ positioning system frame is fully intact after removal, including all components listed (echo 2¿ positioning frame, connectors, suture) and discard appropriately." note: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : sample not returned.
 
Event or Problem Description
As per medwatch form, "patient to the or for ra (operating room for regional anesthesia) laparoscopic incisional umbilical hernia repair with bard 11.4cm ventralight mesh.While patient was in pacu (post-anesthesia care unit,), it was realized that the scaffold was not removed from the mesh during the initial procedure.Patient was taken back to the operating room for a diagnostic laparoscopy.The echo 2 scaffolding was identified and removed with luq (left upper quadrant) trochar.The physician carefully examined the scaffolding and it was determined to be intact.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VENTRALIGHT ST W/ ECHO 2
Common Device Name
SURGICAL MESH
Manufacturer (Section D)
DAVOL INC., SUB. C.R. BARD, INC. -1213643
100 crossings blvd.
warwick RI 02886
Manufacturer (Section G)
BARD SHANNON LIMITED -3005636544
san geronimo industrial park
lot #1, road #3, km 79.7
humacao PR 00791
Manufacturer Contact
andrew topoulos
100 crossings blvd.
warwick, RI 02886
8005566756
MDR Report Key17867824
Report Number1213643-2023-00356
Device Sequence Number5493710
Product Code FTL
UDI-Device Identifier00801741130977
UDI-Public(01)00801741130977
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K170294
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Risk Manager
Type of Report Initial
Report Date (Section B) 09/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/28/2023
Device Catalogue Number5990011
Device Lot NumberHUGT0811
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/12/2023
Initial Report FDA Received Date10/04/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-