Catalog Number 1012534-80
Medical Device Problem Codes Difficult to Remove (1528); Mechanical Jam (2983); Activation Failure (3270)
Health Effect - Clinical Code Unspecified Tissue Injury (4559)
Date of Event 08/14/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the procedure was to treat the superficial femoral artery (sfa).The 6.0x80mm absolute pro vascular self-expanding stent system (sess) would not deploy completely and was stuck within the catheter.Thumbwheel became stuck and would not advance thus making it impossible to complete deployment of the stent.The stent system was removed through the sheath dragged to the common femoral artery (cfa) access site.Stent was pulled out and stuck in the sheath which caused delay in procedure and damage to access site.Surgical cutdown was performed to treat the damaged access site and prolonged hospitalization for access site healing.No additional information was provided.
 
Additional Manufacturer Narrative
Manufacturer's investigation is still pending at this time.Results and conclusions will be provided in the final report.
 
Additional Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.As there was no damage noted to the device during the inspection prior to use, it is possible that during advancement the device was bent in the anatomy resulted in preventing the shaft lumens from moving freely resulting in the reported thumbwheel difficulties and the reported difficulty deploying the stent; however this cannot be confirmed.The investigation was unable to determine a conclusive cause for the reported mechanical jam and the reported activation failure.During removal of the compromised device interaction with the sheath resulted in the reported difficult to remove resulting in a delay in procedure and damage to access site.The treatment appears to be related to the operational context of the procedure as a surgical cutdown was performed to treat the damaged access site and prolonged hospitalization for access site healing.There is no indication of a product quality issue with respect to manufacture, design or labeling.D9/h3 - device available for evaluation updated from ¿yes¿ to ¿no¿.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ABSOLUTE PRO VASCULAR SELF-EXPANDING STENT SYSTEM
Common Device Name
SELF EXPANDING PERIPHERAL STENT SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 2024168
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key17868577
Report Number2024168-2023-10825
Device Sequence Number17564414
Product Code NIP
UDI-Device Identifier08717648176005
UDI-Public08717648176005
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P110028
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2026
Device Catalogue Number1012534-80
Device Lot Number3060861
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/14/2023
Supplement Date Received by Manufacturer12/08/2023
Initial Report FDA Received Date10/04/2023
Supplement Report FDA Received Date12/28/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/08/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexUnknown
-
-