MAUDE Adverse Event Report: MEDTRONIC MEXICO AXIUM PRIME BRPL 3D; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MEXICO AXIUM PRIME BRPL 3D; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION Back to Search Results
Model Number APB-4-12-3D-SS
Medical Device Problem Code Separation Failure (2547)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/26/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
E1.The initial reporter/healthcare professional (hcp) information was not provided in the reported information and it was indicated that the hcp identifying information would not be made available medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may have not been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received report that an axium coil failed to detach intra-operatively.It was unknown if the device was prepared per instructions for use (ifu).There was no harm or injury to the patient.
 
Additional Manufacturer Narrative
H3.Product analysis: equipment used: vis (m-78210), 203cm ruler (m-83361) as found condition (condition of returned device): an axium prime coil was returned for analysis within a shipping box; within a sealed biohazard pouch; within an opened axium prime coil inner pouch and within a resealable bag.The axium implant coil was returned without the introducer sheath.Visual inspection/damage location details: the actuator interface and break indicator were found to be intact.This is indicative mechanical detachment attempts were not made.Upon inspection the axium prime coil pushwire was found to be kinked at ~153.1cm and at ~163.4cm from proximal end.The coin was found not against the lumen stop.The shield coil was found intact with the implant coil still attached.The implant coil was found to be stretched and damaged.No other anomalies were observed.Testing/analysis (including sem reports): the distal outer jacket was removed, and a manual detachment was then attempted by retracting the release wire and the release wire successfully detached implant coil.Conclusion: based on the analysis performed, the customers report of ¿non-detachment¿ was confirmed as the pusher was returned with the implant coil still attached.However, the implant coil was successfully detached manually during testing.The likely cause for the non-detachment are the bends and kinks found on the pusher, causing the release wire to become stuck.The implant coil was found damaged, and the pusher was found bent/kinked.Possible causes for coil and pusher damage are patient vessel tortuosity, device was not hydrated, continuous flush was not administered during procedure, advancing device against resistance or damaged during shipment.H6.Coding updated based on analysis results.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AXIUM PRIME BRPL 3D
Common Device Name
DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION
Manufacturer (Section D)
MEDTRONIC MEXICO
av. paseo del cucapah #10510
tijuana,bc 22570
MX  22570
Manufacturer (Section G)
MEDTRONIC MEXICO
av. paseo del cucapah #10510
tijuana,bc 22570
MX   22570
Manufacturer Contact
glen belmer
9775 toledo way
irvine, CA 92618
6122713209
MDR Report Key17869044
Report Number9612164-2023-04477
Device Sequence Number7028275
Product Code KRD
UDI-Device Identifier00847536008120
UDI-Public00847536008120
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K081465
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/01/2025
Device Model NumberAPB-4-12-3D-SS
Device Catalogue NumberAPB-4-12-3D-SS
Device Lot Number223892158
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/27/2023
Supplement Date Received by Manufacturer11/24/2023
Initial Report FDA Received Date10/04/2023
Supplement Report FDA Received Date12/20/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/01/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-