E1.The initial reporter/healthcare professional (hcp) information was not provided in the reported information and it was indicated that the hcp identifying information would not be made available medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may have not been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
H3.Product analysis: equipment used: vis (m-78210), 203cm ruler (m-83361) as found condition (condition of returned device): an axium prime coil was returned for analysis within a shipping box; within a sealed biohazard pouch; within an opened axium prime coil inner pouch and within a resealable bag.The axium implant coil was returned without the introducer sheath.Visual inspection/damage location details: the actuator interface and break indicator were found to be intact.This is indicative mechanical detachment attempts were not made.Upon inspection the axium prime coil pushwire was found to be kinked at ~153.1cm and at ~163.4cm from proximal end.The coin was found not against the lumen stop.The shield coil was found intact with the implant coil still attached.The implant coil was found to be stretched and damaged.No other anomalies were observed.Testing/analysis (including sem reports): the distal outer jacket was removed, and a manual detachment was then attempted by retracting the release wire and the release wire successfully detached implant coil.Conclusion: based on the analysis performed, the customers report of ¿non-detachment¿ was confirmed as the pusher was returned with the implant coil still attached.However, the implant coil was successfully detached manually during testing.The likely cause for the non-detachment are the bends and kinks found on the pusher, causing the release wire to become stuck.The implant coil was found damaged, and the pusher was found bent/kinked.Possible causes for coil and pusher damage are patient vessel tortuosity, device was not hydrated, continuous flush was not administered during procedure, advancing device against resistance or damaged during shipment.H6.Coding updated based on analysis results.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|